A laboratory rejects a DOT specimen for testing when it finds a fatal flaw, or a correctable flaw that is not corrected in time. Under 49 CFR 40.83(c), the laboratory inspects each specimen and Custody and Control Form (CCF) for these fatal flaws:
Under 40.83(d), the laboratory documents its findings, stops the testing process and reports the result as Rejected for Testing. Under 40.83(h), a broken seal or too little specimen is not fatal if the bottles can be redesignated, for example when the primary bottle leaked but the sealed split bottle holds enough specimen. The laboratory then marks Bottle B as A and Bottle A as B, and initials and dates the change.
Some flaws can be corrected. If the collector's signature is missing from Step 4, the laboratory keeps testing, holds the specimen for at least 5 business days and tries to obtain a correction under 49 CFR 40.205(b)(1). If a non-Federal or expired CCF was used, the laboratory also holds the specimen for at least 5 business days and tries to obtain a correction under 40.205(b)(2). In either case, an uncorrected flaw is reported as Rejected for Testing. If the urine temperature was not checked and the Remarks line has no entry about it, the laboratory tries for five business days to obtain a correction under 49 CFR 40.208 before reporting the result. Under 40.208(b), this omission does not cancel the test. How a Rejected for Testing result affects the test is covered in What Does a Cancelled DOT Drug Test Mean?.