Under 49 CFR 40.235(e), an employer applies the EBT use and care requirements of 49 CFR 40.233 to breath alcohol screening devices (ASDs), and those requirements include external calibration checks at the intervals the manufacturer's instructions specify. The ASD rules in 40.235 apply as well.
Applied to breath ASDs, the requirements of 40.233(c) cover:
Some devices on ODAPC's ASD list work differently. According to ODAPC footnotes, certain models use replaceable semiconductor detectors, and instead of re-calibrating the device, a new calibrated detector can be installed. Another listed system uses a single-use disposable breath tube with an electronic analyzer. The analyzer and tubes are lot specific and manufactured to remain calibrated throughout the shelf life of the device, and the device cannot be used after the expiration date.
Under 49 CFR 40.267, a breath tube ASD test is cancelled if the analyzer has not been pre-calibrated for that device's specific lot, as covered in When Is a DOT Alcohol Test Cancelled?. Calibrating units are covered in What Are Calibrating Units for Breath Alcohol Testers?, and the QAP for ASDs is covered in What Quality Assurance Rules Apply to Alcohol Screening Devices?.